1. Enter the serial number on the pipette handle in the serial number, select the range of the pipette to be calibrated, and select the room temperature. After the input is completed, the measurement volume, accuracy () and precision () of the pipette to be calibrated will appear automatically
Product
Volume Range
Temperature (℃)
Test volume (μl)
Accuracy (%)
Precision (%)
2. In the second step, in the form 1-5, enter the results of 5 consecutive weighings in sequence (unit: mg)
1:
2:
3:
4:
5:
3. Click the "Data Analysis" button to get the test result. Both the accuracy and precision should be within the range of its standard value.
Test results
Average value (μl)
Accuracy (%)
Precision (%)
New calibration
We look forward to cooperation with you